© Bayer MaterialScience AG
Hydrophilic Coatings
Hydrophilic coatings of the BaymedixTM CH series show low swelling in water and a particular hemocompatibility. These coatings are supplied either as aqueous dispersions or in mild solvent solutions, allowing simple dip or spray application processes.
Key Features of Baymedix CH: Hydrophilic Polyurethane Coatings
- Hydrophilic surface: reduced friction, improved biocompatibility
- Strong and durable like traditional polyurethane coatings
- Low-swelling (< 25% by mass)
- Retain mechanical properties when hydrated (see table for details)
- Hemocompatibility exceeds other medical polyurethanes
- Applicable by standard dip- or spray coating processes
| |

|
- 40 % solids in water, dilutable
- solvent-free
- no external surfactants
- low viscosity
(<100 mPa.s / cP)
|
|
- 30 % solids
- mild, low-boiling solvent (toluene/
isopropanol/
methoxypropanol mix)
- high viscosity (>20,000 mPa.s /cP)
- dilutable, as application needs
|
| Product Code | Form | Condition | E-Modulus
MPA (PSI) | Tensile Strength
MPA (PSI) | Elongation
at break % |
.
BaymedixTM
CH 110 | .
.
.
Aqueous
Dispersion | dry, room temperature | 2.6 (370) | 25 (3600) | 550 |
| wet, after 24h in water | 2.5 (360) | 18 (2600) | 450 |
.
BaymedixTM
CH 120 | dry, room temperature | 6.0 (860) | 23 (3300) | 370 |
| wet, after 24h in water | 7.2 (860) | 21 (3000) | 330 |
.
BaymedixTM
CH 310 | .
.
.
Solvent
Solution | dry, room temperature | 2.3 (330) | 25 (3600) | 650 |
| wet, after 24h in water | 2.7 (380) | 27 (3900) | 640 |
.
BaymedixTM
CH 320 | dry, room temperature | 6.0 (860) | 30 (4300) | 420 |
| wet, after 24h in water | 6.5 (920) | 29 (4200) | 440 |